DUBOSCQ, Cristina et al. Estudio del desempeño analítico del dispositivo portátil microINR en pacientes anticoagulados con acenocumarol. Acta bioquím. clín. latinoam. [online]. 2022, vol.56, n.4, pp.481-489. ISSN 0325-2957.

The international normalised ratio (INR) is used to monitor vitamin K antagonist anticoagulated patients. The aim of this study was to evaluate the performance of the microINR portable device in acenocoumarol anticoagulated patients. Group 1: patients provided capillary and venous blood samples for parallel testing comparing microINR with five different pairs of reagent/clot detection systems: brain rabbit thromboplastin with nephelometric, photooptic and viscocity clot detection and recombinant human thromboplastin with nephelometric and photooptical detection. All thromboplastins have an instrument-specific International Sensitivity Index (ISI) lower than 1.10 reported by the manufacturer. Group 2: microINR results were compared with CoaguChek as an established device. Precision was assessed using liquid control materials. The control material coefficient of variation obtained in microINR device was 6.2%. The Passing-Bablok and Bland-Altman regression analysis showed a correlation coefficient greater than 0.92 and a small bias close to zero for all comparisons of microINR with each traditional coagulation method performed on venous blood samples. Both portable devices had a good correlation (r=0.98) and a very low bias of 0.02. The results show clinical agreement of 100% with a Kappa greater than 0.63 for different traditional INR and greater than 0.83 for CoaguChek. The performance microINR is suitable for the anticoagulation control of patients taking acenocumarol.

Keywords : Acenocoumarol; Potable devices; International normalised ratio.