SUELDO, Emanuel; DUBOSCQ, Cristina y ARIAS, Mirta. Validación del ensayo FVIII: C cromogénico en una plataforma automatizada. Acta bioquím. clín. latinoam. [online]. 2020, vol.54, n.2, pp.135-143. ISSN 0325-2957.
The objective of this work was to perform the analytical validation of the chromogenic method (FVIII:Ccro) on the ACL TOP platform correlating with one stage assay (FVIII:Ccoag). The validation study (EP5-A2, EP6-A2 and comparison of methods by EP-9) was performed for the low-normal range curve (CRNB): approximately between 10-150 IU/dL of FVIII and very low range (CRMB): approximately between 0-10 IU/dL. The results of CVr (repeatability) and CVi (intermediate precision) were lower than 6% and comparable to those reported by the manufacturer for other platforms. The analytical measurement range was 11-129 IU/dL, extrapolated to 0.3 IU/dL using the CRMB. For CRNB FVIII:Ccro showed good correlation with FVIII:Ccoag: r: 0.98, slope: 0.982 (0.961-1.003), intercept: -0.3 (-1.1-0.5), bias: -2.0%. For CRMB: r: 0.96 was obtained, pending: 0.921 (0.855-0.988), intercept: -0.07 (-0.35-0.20), bias: -10.2%. Only 4 patients presented discrepant levels between both methods. The automated chromogenic FVIII assay in the ACL TOP family is comparable with one stage coagulable FVIII in the analytical range studied. The FVIII:Ccro automated can be used for the diagnosis and monitoring of the treatment of hemophilic patients.
Palabras clave : Chromogenic FVIII:C assay; Analytical validation methods; Haemophilia A.