Rev. argent. cardiol. vol.86 no.1 Ciudad Autónoma de Buenos Aires feb. 2018

 

ARTICULO ORIGINAL

Usefulness of remote Monitoring of Pediatric Patients with Cardiac implantable electronic devices

Utilidad de los dispositivos cardíaco-eléctricos con monitoreo a distancia en pacientes pediátricos

SEBASTIÁN MALDONADO1 

MARÍA V. LAFUENTE1 

MÓNICA BENJAMÍN1 

EDGARDO GONZÁLEZ PUCHE1 

GISELA DORME, DIEGO LONGARINI2 

MARIANNA GUERCHICOFF2 

PABLO GARCÍA DELUCIS1 

ALBERT O SCIEGATA1 

1 Department of Pediatric Electrophysiology, Hospital de Pediatría SAMIC J. P. Garrahan. Hospital 'Prof. Dr. Juan P. Garrahan

2Department of Pediatric Electrophysiology, Hospital Italiano de Buenos Aires

ABSTRACT:

Background: The aim of this study is to evaluate the usefulness of cardiac implantable electronic devices with remote monitoring system in a pediatric population and the limitations of its implementation in Argentina.

Methods: Twenty-seven patients receiving a cardiac implantable electronic device with remote monitoring system at Hospital Nacional Garrahan and Hospital Italiano de Buenos Aires were included in the study. The rate of events, complications and device-related therapies were evaluated. The anticipated actions taken in response to alert notifications were described. Mean follow-up was 46.6 ± 32.1 months.

results: Median age was 12.2 years (IQR: 8.75-13.3). An implantable cardioverter defibrillator device was placed in 7 patients (25.9%) and 20 (74%) underwent pacemaker implant. Five patients (18.5%) presented seven red alerts: 3 due to ventricular arrhyth-mia in monitoring zone of ventricular fibrillation and 4 due to lead dysfunction. Twelve patients (44%) presented a yellow alert: 6 due to lead dysfunction, 4 due to deactivation of the monitoring system because of lack of signal reception, one due to ventricular tachycardia and another with sinus tachycardia in monitoring zone of ventricular tachycardia. Active actions were taken in 9 pa-tients (33.3%) to manage the alert notification: the atrial lead was replaced in one patient and the ventricular in lead in another; in 2 patients non-compliance with pharmacological treatment and exercise limitation were detected and in the rest of the patients, the device was reprogrammed according to the abnormalities observed in the recording or capture.

Conclusions: Remote monitoring of cardiac implantable electronic devices is very useful in the pediatric population, allowing for the rapid detection and management of device failure or significant arrhythmias.

Key words: Pacemaker; Artificial; Defibrillators; Implantable; Remote Sensing Technology/methods; Monitoring; Physiologic/ methods

RESUMEN:

Objetivo: Evaluar la utilidad de los dispositivos cardíacos eléctricos con sistema de monitoreo a distancia en una población pediátrica y las limitaciones de su implementación en la República Argentina.

Material y métodos: Se incluyeron 27 pacientes a quienes se le implantó un dispositivo cardíaco-eléctrico implantable con sistema de monitoreo a distancia en el Hospital Nacional Garrahan y en el Hospital Italiano de Bs. As. Se evaluó la tasa de eventos, complicaciones y terapias por parte de los dispositivos. Se describieron las conductas anticipadas según la alerta recibida. Se realizó un seguimiento medio de 46,6 meses ± 32,1.

resultados: La edad fue 12,2 años (RIC: 8,75-13,3), a 7 pacientes (28%) se les implantó un cardiodesfibrilador implantable y 20 pacientes (78%) un marcapaso endocavitario. Cinco pacientes (18,5%) presentaron 7 alertas rojas: 3 por arritmia ventricular en rango de fibrilación ventricular y 4 por alteraciones en alguno de los cables. Doce pacientes (44%) presentaron una alerta amarilla: 6, por alteraciones en los cables; 4, por desactivación del sistema por falta de recepción de señal; 1, por taquicardia ventricular; y 1, por taquicardia sinusal en rango de taquicardia ventricular. En 9 pacientes (33,3%) se tomó una conducta activa para resolver el aviso de alerta: en 2 pacientes se realizó recambio de cable auricular en uno y ventricular en otro, en 2 pacientes se detectó incumplimiento del tratamiento farmacológico y en la limitación del ejercicio, en el resto se reprogramó el dispositivo, según el tipo de alteración en el registro o la captura.

Conclusiones: El sistema de monitoreo remoto de los dispositivos cardíaco-eléctricos implantables es muy útil en la población pediátrica, lo que permite una rápida detección y acción cuando se produce un fallo en el dispositivo o un evento arrítmico de relevancia clínica.

Palabras claves: Marcapasos artificial; Desfibriladores implantables; Tecnología de sensores remotos/métodos; Monitoreo fisiológico/métodos

INTRODUCTION

Over the past 30 years, the technological development of cardiac implantable electronic devices (CIED) has increased, allowing better programming to reduce complications and increase benefits. In 1971, the first remote CIED monitoring via telephone connections was launched and, since then, with the advent of the Internet, remote interrogation of the devices has im-proved as well as the access to the data stored regard-ing device performance or the presence of arrhythmi-as during follow-up. (1) Remote CIED monitoring is a safe strategy that lowers the number of ambulatory visits and reduces costs for the health care system and patients, especially in those countries with central-ized systems with extensive territorial surface areas such as Argentina. (2) On the other hand, it provides the possibility of taking actions before the follow-up visit, reducing complications and unnecessary visits to the emergency room, and permits the indication of preventive treatments despite the distance. (3) In the pediatric population, CIED implant presents distinctive features that makes it different from adults, such as growth, increased inflammatory reaction and fibro-sis, greater need of battery replacement and, in the case of congenital heart defects, unfavorable anatomy for device insertion. All these factors explain the high-er incidence of complications associated with CIED implantation in children. (4) Many publications have reported the benefits of remote monitoring systems, particularly in the adult population. (3, 4) In pediat-ric patients, the evidence is lower but the incidence of complications is higher. Therefore, the use of this system would allow physicians to detect these com-plications earlier, reducing risks, particularly in those patients living in areas distant from health care centers with capabilities for evaluating CIED recipients. The aim of our study was to evaluate the usefulness and feasibility of this monitoring system in a pediatric population belonging to two tertiary care hospitals with national coverages.

METHODS

This retrospective and observational cohort study included consecutive patients undergoing implantation of pacemakers or implantable cardioverter defibrillator (ICD) devices with remote monitoring system. The study was conducted in two medical centers: Hospital Nacional J. P Garrahan and Hospital Italiano de Buenos Aires. The clinical characteristics of the population, the rate of events, complications and therapies delivered by the devices were evaluated in the long-term. Actions taken in response to alert notifications were described, as well as the difficulties of the system to transmit the information. The follow-up period was 46.6± 32.1 months.

IQR Interquartile range VT Ventricular tachycardia

Patients who lived within 100 km from the center were invited to be examined every 6-8 months, depending on their preferences, while those who lived further away could be examined every 8 to 12 months in the absence of alert notifi-cations. In case of a red alert, the clinical staff immediately called the patient’s relatives and indicated the actions to be taken according to the findings. In case of a yellow alert, the clinical implication of the notification was evaluated and the patient was contacted if it was necessary.

Statistical analysis

Data were stored using Microsoft Office Excel 2010. All the calculations were performed using Stata 12.0 software statistical package.

Frequency and/or percentage distribution was estab-lished for all the variables in relation with the total number of cases and the values were expressed as proportions, mean and standard deviation (SD) or median and interquartile range (IQR), as applicable.

ethical considerations

The study was conducted following the recommendations of the 1975 Declaration of Helsinki, corrected in 1983 and revised in1989, and was approved by the Hospital Garrahan ethics committee.

RESULTS

Population characteristics

The cohort consisted of 27 patients. Median age was 12.2 years (IQR: 8.75-13.3) and 11 patients (40.5%) were boys. Figure 1 shows the geographical distribu-tion of the patients, with 85.1% (23 patients) living in regions more than 100 km away from the health care center. Nine patients (33.3%) presented congenital heart defects. Seven patients (25.9%) received an ICD device and 20 (74%) underwent pacemaker im-plantation. The pacemakers were implanted due to complete AV block in 17 (62.9%) patients and to sick sinus syndrome in 3 (11.1%). Implantable cardiovert-er defibrillators were implanted for secondary preven-tion in all the patients, except in one. Its indication was due to channelopathies in 5 (18.5%) patients and to ventricular tachycardia (VT) in 2 (7.4%) patients with congenital heart defects. Twenty-five Biotronik devices (Home Monitoring System) and two Abbot-St Jude Medical devices (Merlin Home) were implanted.

analysis of remote Cied monitoring

Five patients (18.5%) presented seven red alerts. Three alerts were due to tachycardia in the monitoring zone of ventricular fibrillation; in two of these alerts, patients received appropriate shocks and the episode reverted spontaneously in one case; one patient presented a sudden change first in the atrial lead impedance and then in the ventricular lead; in one patient the alert was due to insulation failure of the atrial lead with loss of capture at maximal output; another patient presented fracture of the atrial lead, and in the last alert the patient received an inappropriate shock to due noise in a fractured ventricular lead. Mean time from the activation of the monitoring system to the event was 23.8 ± 14.9 months.

Fig. 1 a. Place of residence of the patients with cardiac implantable electronic devices with remote monitoring system. B. Geographic location of the cities in which the patients live. 

Twelve patients (44%) presented a yellow alert: 6 due to lead dysfunction (2 with high capture thresh-olds, 2 with recordings below the cutoff limit established and 2 with atrial lead noise), 4 due to deactivation of the remote monitoring system because of lack of signal reception, and 2 due to arrhythmias (one with VT that reverted with antitachycardia pacing therapy and one patient with an ICD that presented several episodes of sinus tachycardia in monitoring VT-1 zone). Mean time from the activation of the monitoring system to the yellow alert was 17.5 ± 15.7 months. Figure 1 shows some examples of the alert notifications received. analysis of the actions taken according to the findings

Active actions were taken in 9 patients (33.3%) to manage the alert notification: the atrial lead was re-placed in one patient and the ventricular lead in an-other; in 2 patients in whom non-compliance with pharmacological treatment and exercise limitation

was detected, counseling on therapeutic measures was intensified and antiarrhythmic treatment was adjusted; in the rest of the patients, the device was repro-grammed according to the abnormalities observed in the recording or capture. Temporary deactivation of the system due to lack of signal occurred in 4 patients (14.8%) and was solved by relocating the receiver. The signal could not be reactivated in only one patient living in the jungle of Misiones province, a region where the Wi-Fi signal is very weak. During follow-up, no patient indicated that they would like to discontinue the use of the remote monitoring system.

DISCUSSION

Our registry of pediatric patients with CIED is the first one performed in our country. The study showed a high incidence of clinical events in pediatric pa-tients, including significant arrhythmias as V T, ven-tricular fibrillation and atrial flutter, lead dysfunction (recording and capture) and lead integrity abnormali-ties (impedance, change of polarity). While the study by Malloy et al. reported that arrhythmias (atrial fi-brillation in 70% of the cases) and lead abnormalities were the most common events during follow-up, the most prevalent findings in our study were lead dysfunction and ventricular tachycardia/fibrillation. (5) This difference can be explained by the age of the patients, that was 21 years in the study by Malloy et al. and 12.2 years in ours. Leoni et al. evaluated a population of pediatric patients with an age similar to that of our study (9.4 years) and reported lead dysfunction as the most common clinical event. (6) The incidence of arrhythmias associated with congenital heart defects and some channelopathies increases with age, while lead dysfunction is more common in younger children and is related with growth and heart size. (7, 8) In 75% of the alert notifications an action was taken before the scheduled hospital visit, preventing the develop-ment of complications, as shown in the ECOST study in patients with ICD, where inappropriate shocks were reduced by 71% and capacitor charges by 76%, increasing battery longevity. (9) Also, the CONNECT study, conducted in patients implanted with an ICD or a cardiac resynchronization therapy device, showed an 80% reduction from the time an event occurred to clinical decision, while this interval evidenced 25% reduction in patients with pacemaker evaluated in the PREFER study. (10, 11) The use of remote CIED monitoring allows remote management and reduces face-to-face visits without increasing morbidity and mortality as reported in the TRUST study, where remote monitoring reduced the number of in-hospital visits by 45% without presenting major complica-tions. (3) During follow-up, a high rate of deactivation was observed mainly due to lack of the Wi-Fi signal. In most cases the problem was solved by relocating the receiver inside the patient’s home. Yet, this finding represents a potential limitation to take into ac-count when indicating this system to a patient, since

the quality and distribution of the Wi-Fi signal is very heterogeneous in our country and this may complicate its implementation in some regions. Before indicating this type of system, it would be better to evaluate the environment in which the patient lives and the qual-ity of the Wi-Fi signal to avoid implanting an expen-sive device which will not be able to be used. Despite these devices are far more expensive than convention-al ones, several publications have demonstrated that anticipated actions guided by the remote monitoring system avoid complications and unnecessary in-office visits which are more costly than remote CIED moni-toring, particularly if the costs of transportation of patients who live in regions far away from the health care centers are included as a variable. (12, 13) in the cost-benefit relationship.

Study limitations

The results are based on the retrospective analysis of data.

CONCLUSIONS

Remote CIED monitoring is useful in the pediatric population as it allows rapid detection and manage-ment of device failure or clinically significant arrhyth-mias. This benefit is even greater in those patients living far away from the hospital, allowing for longer time intervals between visits.

REFERENCES

1. Varma N, Piettro Ricci R. Impact of remoting monitoring on clini-cal outcome. J Cardiovasc Electrophysiol 2015; 26: 1388-95. http:// doi.org/f752xgLinks ]

2. Mabo P, Victor F, Bazin P, Ahres S, Babuty D, DaCosta A, et al; COMPAS Trial Investigators. A randomized trial of long-term remote monitoring of pacemaker recipients. Eur Heart J 2012;33:1105-11. http://doi.org/cj89tgLinks ]

3. Varma N, Michalski J, Epstein AE, Schweikert R. Automatic remote monitoring of implantable cardioverter-defibrillator lead and generator performance: the Lumos-T Safely RedUceS RouTine Office Device Follow-up (TRUST)Trial. Circ Arrhythm Electrophysiol 2010;3:428-36. http://doi.org/d7hvn5Links ]

4. Czosek RJ, Meganathan K, Anderson JB, Knilans TK, Marino BS, Heaton PC. Cardiac rhythm devices in the pediatric population: Utilization and complications. Heart Rhythm 2012;9:199-208. http:// doi.org/fnbcvpLinks ]

5. Maloy L, Gingerich J, Olson M, Atkins D. Remote Monitoring of Cardiovascular Implantable Devices in the Pediatric Population Improves Detection of Adverse Events. Pediatr Cardiol 2014;35:301-6. http://doi.org/f5trjjLinks ]

6. Leoni L, Padalino M, Biffanti R, Ferretto S, Vettor G, Corrado D, et al. Pacemaker Remote Monitoring in the Pediatric Population: Is It a Real Solution? Circulation 1990;81:1544-9. [ Links ]

7. Matsuhisa H, Oshima Y, Maruo A, Hasegawa T, Tanaka A, Noda R, et al. Pacing therapy in children. Circ J 2014;78:2972-8. http:// doi.org/cmskLinks ]

8. Wilhelm BJ, Thöne M, El-Scheich T, Livert D, Angelico R, Osswald B. Complications and Risk Assessment of 25 Years in Pediatric Pac-ing. Ann Thorac Surg 2015;100:147-53. http://doi.org/f7m6jhLinks ]

9. Guedon-Moreau L, Lacroix D, Sadoul N, Clementy J, Kouakam C, Hermida JS, et al. A randomized study of remote follow-up of im-plantable cardioverter defibrillators: safety and efficacy report of the ECOST trial. Eur Heart J 2012;34:605-14. http://doi.org/f4qpgxLinks ]

10. Crossley G, Boyle A, Vitense H, Chang Y, Mead RH. The CON-NECT (Clinical Evaluation of Remote Notification to Reduce Time to Clinical Decision) trial: the value of Wireless remote monitoring with automatic clinician alerts. J Am Coll Cardiol 2011;57:1181-9. http://doi.org/fpcggxLinks ]

11. Crossley GH, Chen J, Choucair W, Cohen TJ, Gohn DC, Johnson WB, et al. Clinical benefits of remote versus transtelephonic monitoring of implanted pacemakers. J Am Coll Cardiol 2009;54:2012-9. http://doi.org/d966z9Links ]

12. Landolina M, Perego GB, Lunati M, Curnis A, Guenzati G, Vicentini A, et al. Remote monitoring reduces healthcare use and im-proves quality of care in heart failure patients with implantable de-fibrillators: the evolution of management strategies of heart failure patients with implantable defibrillators (EVOLVO) study. Circulation 2012;125: 2985-92. http://doi.org/f35773Links ]

Creative Commons License This is an open-access article distributed under the terms of the Creative Commons Attribution License