Pap. Phys. vol.6 no.1 La Plata ene./jun. 2014

http://dx.doi.org/10.4279/PIP.060001 

ARTICULOS

Experimental determination of L X-ray fluorescence cross sections for elements with 45 ≤ Z ≤ 50 at 10 keV

 

E. V. Bonzi,1,2* G. B. Grad,1 R. A. Barrea3

* E-mail: bonzie@famaf.unc.edu.ar
† E-mail: grad@famaf.unc.edu.ar
‡ E-mail: rbarrea@depaul.edu

1 Facultad de Matemática, Astronomía y Física, Universidad Nacional de Cordoba. Ciudad Universitaria. 5000 Córdoba, Argentina.
2 Instituto de Física Enrique Gaviola (CONICET), 5000 Cordoba, Argentina.
3 Physics Department, DePaul University, Chicago, IL 60614, USA.


Received: 19 December 2013, Accepted: 20 January 2014
Edited by: P. Weck
Reviewed by: J. P. Marques, Departamento de Física, Centro de Física Atómica, Fac. de Ciencias, Universidade de Lisboa, Portugal.

Licence: Creative Commons Attribution 3.0 DOI: http://dx.doi.org/10.4279/PIP.060001


Synchrotron radiation at 10 keV was used to experimentally determine the LI, Lα, LβI, LβII, LγI and LγI I fluorescence cross sections for elements with 45 ≤ Z ≤ 50, as part of an ongoing investigation at low energies. The measured data were compared with calculated values obtained using coefficients from Scofield, Krause and Puri et al.

 

I. Introduction

This work is part of a systematic investigation on elements with 45 ≤ Z ≤ 50, which has been carried out at different energies [1-3]. The L X-ray cross sections were measured with monoenergetic excitation beam at 10 keV.

We report cross sections for each spectral line, according to the resolution of the Si(Li) solid state detector used to resolve individual component lines of the spectral emission. The experimental cross sections were grouped considering the transitions scheme, the energy of the emission lines and the detector resolution.

In general, the fluorescence cross sections obtained in this work show the same trend with Z and broad agreement with the data published by Puri et al. [4,5] and Krause [6,7], calculated using Scofield's coefficients [8,9].

II. Experimental Condition

The measurements were carried out at the X-ray Fluorescence beam line at the National Synchrotron Light Laboratory (LNLS), Campinas, Brazil [10]. The components of the experimental setup were:

• Silicon (111) channel cut double crystal monochromator, which can tune energies between 3 and 30 keV. The energy resolution is 3-10-4 to 4-10-4 between 7 and 10 keV.

• A Si(Li) solid state detector, 5 mm thick and 5 mm in diameter, with a resolution of 170 eV at 5.9 keV and a 0.0127 cm thick beryllium window. The model introduced by Jaklevic and Giauque [11] was used to obtain the detector efficiency.

• The whole setup is mounted on a motorized lift table, which allows the vertical positioning of the instruments within the linearly polarized part of the beam.

• To limit the beam size, a motorized computer controlled set of vertical and horizontal slits (located upstream and downstream of the monochromator) was used.

A set of foil samples (rhodium, palladium, silver, cadmium, indium and tin ) was used to determine the L fluorescence cross sections of these elements. The foil samples were provided by Alfa products Inc., with a certified purity of over 99%. The foils thicknesses are shown in Bonzi et al. (see Table I) [2].

Table 1: Experimental and theoretical L X-ray fluorescence cross sections in Barns/atom at 10 keV. Experimental data (This work), theoretical values calculated using Scofield [8] and Puri [4] and semi-empirical coefficients obtained from Scofield [8] and Krause [6].

K emission lines of chlorine, calcium, titanium and iron were measured to determine the geometrical and the detector efficiency factors.

The and Lα fluorescent spectra were measured by collecting 2-105 net counts for each element in order to have the same statistical counting error in all measured spectra.

A system dead time, lower than 1%, was established measuring the fluorescence emission of a Ti sample, adjusting the slit at the exit of the monochromator. All samples were measured with the same slit aperture. Unwanted effects, such as piling up, were avoided using this configuration and the geometric factors were ensured to be the same for all samples. This configuration made it unnecessary to carry out corrections for count losses, spectra distortions or modification of the geometrical arrangement.

III. Spectra analysis

The energy of the emission lines tabulated by Scofield [8,9] and the detector resolution were considered to group the L X-ray fluorescence lines. This line arrangement was used to fit the L spectrum, where the LP and L7 compound lines have been noted with a Roman subscript according to the most intense contribution line, with its corresponding atomic transition:

The background radiation was fitted using a linear second order polynomial.

The area of the fluorescence peaks was determined as the average of the areas obtained by the adjustment using Hypermet and Gaussian functions. The escape peaks were fitted using a Gaussian function.

As a consequence of the excitation with a linearly polarized photon beam, the contribution to the background was very low. The linear polarization of the incident beam produces negligible scattered radiation at 90° with respect to the incident beam direction. The detector position is localized at the same height of the storage ring.

IV. Data Analysis

The expression for the L experimental fluorescence cross sections is [13]

T(Eo,EKi) = correction factor for self absorption in the sample, Eo = energy of the incident beam and EKi = energy of the K spectral line for a given element, from Scofield [8].

The factor Io.G.e(E) was previously determined in Ref. [2], where the same geometry and detector were used. Because of this, the Io.G.e(E) energy dependence is already known and only a scale factor is needed to obtain the correct beam intensity.

Four targets: Cl (NaCl), Ca (CaHPO4 • 2H2O), Ti (Ti foil) and Fe (Fe foil) emitting fluorescent X-rays in the range from 2.4 keV to 7.0 keV were used to determine the scale factor in this work. Four and four Kβ lines were used to fit the scale factor. Jaklevic and Giauque's [11] model was used to fit the detector efficiency.

V. Results and Discussion

L X-ray cross section values obtained in our fluorescence experiment and the theoretical values calculated by using coefficients given by Scofield [8,9], Puri et al. [4] and Krause [7] are shown in Table 1 and Figs. 1 to 6.

Puri et al. predicted theoretical Coster Kronig and fluorescence values using ab initio relativistic calculations, while Krause's values of and were obtained by fitting experimental and theoretical compiled data. In Krause's tables, the theoretical data were calculated for singly ionized free atoms while the experimental data contain contributions from solid state, chemical and multiple ionization effects.

The La cross section values have a better agreement with the theoretical values when the intensity peaks are fitted with an Hipermet function instead of a Gaussian function. This happens because the Hipermet function has a tail on the left side that increases the fitted area.

Moreover, the tail of the Hipermet function used to fit the La peaks diminishes the area and the cross sections of the LI peaks, accordingly.

The experimental LI cross sections show a similar Z trend, compared to the data obtained using Krause and Puri et al. values. Nevertheless,in general, our results are lower than those.

The Lα experimental fluorescence cross section data, Fig. 2, agree well with Krause's values although for elements with higher Z, the experimental values are slightly lower than those from Krause. They are even lower than Puri's et al. values, but still showing the same trend with Z.

The I experimental fluorescence cross sections show a very good agreement with the theoretical values when the Hipermet function is used to fit the area (see Fig. 3).

TheII measured cross sections show a similar dependence on Z as both theoretical assemblies (Fig. 4).

TheI Sn experimental value is lower than the data presented by either Puri et al. or Krause and theII Sn experimental value is much higher than both theoretical values. This behavior might be due to the fitting process as both spectra lines are too close in energy; theII intensity seems to be overestimated while the I intensity seems to be underestimated.

A similar behavior is observed for rhodium experimental data although the differences with the theoretical values are much smaller than those for tin.

The experimental I fluorescence cross sections show some differences with the Z trend of the theoretical data: in the lower Z range, the experimental values are higher than the theoretical ones while for higher Z, this difference becomes smaller. Sn values, Z = 50, show a different behavior, being lower than both calculated values (see Fig. 5).

The Ly// experimental values show the general Z trend of the values presented by Krause and Puri et al. . The experimental values are sometimes higher or lower than the theoretical ones but the range of values is similar to them (see Fig. 6).

To determine the uncertainties of the experimental cross sections, the propagation of errors was carried out in Eq. (1). The uncertainty values are in general around 6-10%, and less than 40% in case of the LI line.

The uncertainty associated to the Io.G.e(E) factor was estimated as the mean quadratic deviation of the experimental values (≤ 2%). For the factor T(Eo, ELi), a propagation of errors was carried out assuming a 3% error in the values of the mass absorption coefficients, and a 2% error in the sine of the angles due to the sample positioning errors. Krause's ujk values for elements with 45 ≤ Z ≤ 50 have an estimated error of 1%.

The uncertainties of the peak areas were established as half the difference between the areas obtained using Gaussian and Hypermet functions to fit. These uncertainties were the main contribution to the experimental errors of the cross section.

VI. Conclusions

In this investigation, the L X-ray fluorescence cross sections of a group of elements with 45 ≤ Z ≤ 50 were measured using a synchrotron radiation source for monoenergetic beams at 10 keV. The polarization properties of the monoenergetic excitation beam and the high resolution of the detector system allowed to reduce the scattered radiation thus obtaining a better signal to noise ratio and a better accuracy for the experimental cross sections.

The cross sections of LI, Lα, I,II, I and Ly// Ilines were measured considering a more detailed group than the usual sets. In Table 1, the comparison between the experimental fluorescence cross section values with the theoretical values calculated using coefficients from Scofield [8,9], Puri et al. [4] and Krause [6] are shown.

Our experimental values are in general in good agreement with the calculated data using Scofield's [8,9] and Krause's [6] coefficients.

The L cross sections present uncertainties around 6-10% and the less intensive LI peaks show uncertainties that in some cases come close to 40%, being the fitting uncertainty the most important error source.

The use of the Hypermet function is very convenient to fit the Lα and peaks (see Table 1).

The solid state detector used in our experiments does not have enough energy resolution to resolve each spectral line. A higher resolution detection system would be desirable in order to analyze each spectral line separately.

The Coster Kronig coefficients present large fluctuations in this atomic range and that is the cause of the observed discrepancies.

Acknowledgements - This work was carried out under grants provided by SeCyT U.N.C. (Argentina). Research partially supported by LNLS -National Synchrotron Light Laboratory, Brazil.

1. E V Bonzi, R A Barrea, Experimental L X-ray fluorescence cross sections for elements with 45 ≤ Z ≤ 50 at 7 keV by synchrotron radiation photoionization, X-ray Spectrom. 34, 253 (2005).

2. E V Bonzi, N M Badiger, G B Grad, R A Barrea, R G Figueroa, Measurement of L X-ray luorescence cross sections for elements with 45 ≤ Z ≤ 50 using synchrotron radiation at 8 keV, Nucl. Instrum. Meth. B 269, 2084 (2011).

3. E V Bonzi, N M Badiger, G B Grad, R A Barrea, R G Figueroa, L X-ray luorescence cross sections experimentally determined for elements with 45 ≤ Z ≤ 50 at 9 keV, Appl. Radiat. Isotopes 70, 632 (2012).

4. S Puri, D Mehta, B Chand, N Singh, P N Tre-han, L shell luorescence yields and costerkro-nig transition probabilities for the elements with 25 ≤ Z ≤ 96, X-ray Spectrom. 22, 358 (1993).

5. S Puri, B Chand, D Mehta, M L Garg, S Nir-mal, P N Trehan, K and L shell X-ray fluorescence cross sections, Atom. Data Nucl. Data 61, 289 (1995).

6. M O Krause, Atomic radiative and radiation-less yields for K and L shells, J. Phys. Chem. Ref. Data 8, 307 (1979).

7. M O Krause, C W Nestor, C J Sparks, E Ricci, X-ray fluorescence cross sections for K and L-rays of the elements, Oak Ridge National Laboratory, Report 5399 (1978).

8. J H Scofield, Theoretical photoionization cross sections from 1 to 1500 keV, Lawrence Liv-ermore National Laboratory, Report 51326 (1973).

9. J H Scofield, Relativistic Hartree Slater values for K and L X-ray emission rates, Atom. Data Nucl. Data 14, 121 (1974).

10. C A Perez, M Radtke, H Tolentino, F C Vicentin, R T Neuenshwander, B Brag, H J Sanchez, M Rubio, M I S Bueno, I M Raimundo, J R Rohwedder, Synchrotron radiation X-ray fluorescence at the LNLS: Beam-line instrumentation and experiments, X-ray Spectrom. 28, 320 (1999).

11. J M Jaklevic, R D Giauque, Handbook of X-ray spectrometry: Methods and techniques, Eds. R Van Grieken, A Markowicz, Marcel Dekker, New York (1993).

12. M C Leypy, J Plagnard, P Stemmler, G Ban, L Beck, P Dhez, Si(Li) detector efficiency and peak shape calibration in the low energy range using synchrotron radiation, X-ray Spectrom. 26, 195 (1997).

13. D V Rao, R Cesareo, G E Gigante, L X-ray fluorescence cross sections of heavy elements excited by 15.20, 16.02, 23.62 and 24.68 keV photons, Nucl. Instrum. Meth. 83, 31 (1993).

14. J H Hubbell, S M Seltzer, Tables ofX-ray mass attenuation coefficients and mass-energy absorption coefficients from 1 KeV to 20 MeV for elements Z=1 to 92 and 48 additional substances ofdosimetric interest, NISTIR, Report 5632 (1995).

15. Md R Khan, M Kαrimi, Ki/Kα ratios in energy dispersive X-ray emission analysis, X-ray Spectrom. 9, 32 (1980).