Titanium alloy orthodontic mini-implants: scanning electron microscopic and metallographic analyses

 

Paola F. P. Burmann1, Henrique C. Ruschel2,4, Ivana A. Vargas1, José C. K. de Verney3, Paulo F. Kramer4

1 Department of Orthodontics. Lutheran University of Brazil (ULBRA), Canoas, RS, Brazil.
2 Department of Oral Histology. Lutheran University of Brazil (ULBRA), Canoas, RS, Brazil.
3 Department of Engineering. Lutheran University of Brazil (ULBRA), Canoas, RS, Brazil.
4 Department of Pediatric Dentistry. Lutheran University of Brazil (ULBRA), Canoas, RS, Brazil.

CORRESPONDENCE Dr. Paola Flach Perim Burmann Rua Antunes Ribas, 2016, Meller Sul Santo Angelo – RS – Brazil Phone/Fax: +55 55 9101 470 e-mail: ortopfperim@hotmail.com


ABSTRACT

Anchorage control is one of the determining factors of successful orthodontic mechanics. In mini-implants, fractures due to placement and removal have been related to implant design and titanium alloy quality. This study assessed the topography and microstructure of five brands of mini-implants (Neodent, SIN, Morelli, Conexao, Foresta Dent). Scanning electron microscopic analyses of the head, transmucosal neck, threaded body, and tip were performed to assess implant design and manufacturing defects (n=3/group). Metallographic analysis of longitudinal sections (n=15) and cross-sections (n=15) was performed under conventional light microscopy according to international standards of "American Society for Testing and Materials". The results showed significant differences in miniimplant design. Surface irregularities in the threaded body and tip were observed. Microstructural analyses revealed an alpha/beta-phase grain structure, in compliance with the ETTC-2 ("Technical Committee of European Titanium Producers" –2nd edition). No structural defects were detected. We conclude that differences in mini-implant design and the presence of surface irregularities may influence the effectiveness of orthodontic anchorage.

Key words: Orthodontics; Dental implantation; Titanium.

RESUMO

Mini-implantes ortodônticos de liga de titânio: análises ao microscópio eletrônico de varredura e metalográfica

O controle da ancoragem e um dos fatores decisivos no sucesso da mecanica ortodontica. Fraturas devido ao estresse de insercao e remocao de mini-implantes sao associadas ao design das pecas e a qualidade da liga de titanio. O presente estudo analisou a topografia e a microestrutura de cinco marcas de mini-implantes (Neodent, SIN, Morelli, Conexao, Foresta Dent). Analise ao microscopio eletronico de varredura da cabeca e perfil transmucoso, porcao rosqueavel e ponta ativa foi realizada com o proposito de avaliar o design e defeitos de fabricacao (n=3/grupo). A analise metalografica baseou-se nas normas internacionais da "American Society for Testing and Materials" e revelou a microestrutura em cortes longitudinais (n=15) e transversais (n=15) por meio do microscopio optico. Os resultados demostraram que os mini-implantes apresentam diferencas significativas no design. Irregularidades superficiais na porcao rosqueavel e na ponta ativa foram tambem observadas. A analise da microestrutura revelou uma estrutura de graos fases alfa e beta distribuidas de acordo com os padroes definidos pelas normas ETTC-2 ("Technical Committee of European Titanium Producers" – 2a edicao). Alem disso, nao foram detectados defeitos na estrutura interna das ligas. Conclui-se que diferencas no design dos mini-implantes e a presenca de irregularidades superficiais podem influenciar na efetividade da ancoragem durante o tratamento ortodontico.

Palavras-chave: Ortodontia; Implante dentario; Titanio.


 

INTRODUCTION

Orthodontics is based on the exertion and control of forces acting on the teeth and supporting structures. Therefore, control of anchorage is essential for the success of orthodontic treatment1. Traditional orthodontic anchorage depends on patient compliance. Furthermore, the number or quality of teeth is often insufficient for effective anchorage2. Therefore, several anchorage devices have been used in recent decades. Prosthetic implants, plates, and onplants have been replaced with mini-implants because they eliminate the need for invasive surgical procedures, high cost, placement site limitations, and considerable time for osseointegration3. Mini-implants are popular because of their ease of insertion and removal, less discomfort for patients, possibility of immediate loading, high versatility, and low cost4-6. Clinical and laboratory outcomes, however, have shown failure rates of 10 to 30%, mostly related to inflammation of peri-implant tissues, characteristics of soft tissues, and miniimplant placement site7-9. Screw diameter, length, thread form, presence of flutes, and screw material have also been implicated in poor primary stability of these devices10-13.
The optimal material of mini-implants would exhibit excellent corrosion resistance, biocompatibility, and sufficient mechanical strength to resist placement and removal14. Titanium alloys have been used in these devices. The use of vanadium and aluminum have significantly enhanced their performance and mechanical properties15. Nevertheless, studies on the internal microstructure of miniimplants rare in the literature14-16.
Because of the diameter and length restrictions of mini-implants, optimal shape design is important for primary stability. The strength resistance of a titanium alloy depends on its microstructure, which is influenced by the composition, heat treatment, and machining processes of the mini-implant17. Thus, studies analyzing the topography and microstructure of mini-implants are essentially important. The objective of the present study was to analyze the topographical and microstructural features of miniimplants used for orthodontic anchorage.

MATERIALS AND METHODS

Topography and microstructure were analyzed on Ti-6Al-4V (Grade 5 titanium alloy) self-drilling orthodontic mini-implants from five different dental implant manufacturing companies (four Brazilian and one imported). They were allocated into five groups: Group 1 – NeodentR (Curitiba, Parana, Brazil); Group 2 – SINR (Sao Paulo, Sao Paulo, Brazil); Group 3 – MorelliR (Sorocaba, Sao Paulo, Brazil); Group 4 – ConexaoR (Aruja, Sao Paulo, Brazil); and Group 5 – Foresta DentR (Pforzhein, Baden-Wurttemberg, Germany). Scanning electron microscopy (SEM) analysis was conducted to obtain a descriptive analysis of implant design and detect potential manufacturing defects. Three mini-implants of each of the five brands were analyzed. The implants were bonded to aluminum stubs (Sigma Chemical Co., St. Louis, Missouri, USA) with cyanoacrylate adhesive (Super Bonder Gel, Loctite, Diadema, Sao Paulo, Brazil) and immediately analyzed under high-vacuum SEM (Philips XL 20, FEI, Eindhoven, The Netherlands) at 20kV accelerating voltage. Images of the screw head, transmucosal neck, threaded body, and tip were analyzed at 50x, 100x, and 200x magnification. Metallographic analysis was conducted to detect discontinuities and to assess the presence of alpha- and beta-phase titanium. The methodology was based on the "American Society for Testing and Materials" (ASTM International). The standards applied were ASTM E3-01 (Standard Guide for Preparation of Metallographic Specimens)18, ASTM E407-99 (Standard Practice for Microetching Metals and Alloys)19, and ASTM E7-03 (Standard Terminology Relating to Metallography)20. Six mini-implants (three in longitudinal section, three in cross section) from each of the five brands were analyzed. For longitudinal sections, the samples were coldembedded in methyl methacrylate polymer before being sectioned lengthwise. The mini-implants were ground and polished with a series of silicon carbide abrasive sheets - 220, 320, 400, 600, and 1200 grit (3M, Sumare, Sao Paulo, Brazil) under water. The samples were polished in a DP-10 sander (Panambra, Sao Paulo, Sao Paulo, Brazil), using diamond abrasive compound (3M, Brazil) with 1- and 2-mm grains. The longitudinally sectioned and polished mini-implants were etched in a solution of 10mL HF, 5mL HNO3, and 85mL H2O (Kroll's reagent) for 20 seconds, dried with hot air, and analyzed under light microscopy (Union MC 85800, OptiTec Ltd., Japan) at x50 and x400 magnification.
Cross sections were obtained at TORK(Controle Tecnologico de Materiais Ltda, Sao Paulo, Sao Paulo, Brazil). Samples were cold-embedded in acrylic resin. Metallographic analysis was based on the ISO 5832-3 standard (Implants for surgery - Metallic materials - Part 3), and the alpha/beta phases were compared with the European Technical standards (ETTC-2) published by the Technical Committee of European Titanium Producers. The mini-implants were sectioned with a circular table saw (Arotec, Sao Paulo, Sao Paulo, Brazil) and a cutting disc (Norton, Worcester, Massachusetts, USA). The specimens were sanded in a round device with silicon carbide abrasive sheets - 150, 220, 320, 400, and 600 grit (3M, Brazil), on a tabletop grinding machine, using water as a lubricant to obtain a flat, homogeneous surface. The specimens were then polished in a sander with 6- μm and 3-μm diamond abrasive compound and buffed with 1-μm diamond compound. The cross-sectioned specimens were etched with a solution composed of 6 g NaOH, 60 mL H2O, and 10 mL H2O2 for 20 seconds and dried with hot air. This process revealed the microstructure of the mini-implants, in which an effective contrast between alpha and beta phases was observed. The cross sections were examined under light microscopy (OPTON / TNM-07 PL, Cotia, Sao Paulo, Brazil) at x200 magnification.

RESULTS

The mini-implants exhibited significant differences between brands in screw head and transmucosal neck; pitch and shape of threads; and active tip design (Fig. 1). Surface irregularities can result from machining process, polishing defects, crystal growth deposits, and areas of detritus. The greatest amount of surface irregularities and detritus was found on the tips of Groups 1, 2, and 3. The best surface finish along the threaded body was found in Groups 1 and 5. Mini-implants in all groups had adequate surface finish and no evidence of irregularities on the screw head or transmucosal neck (Fig. 1).


Fig. 1
: Head and transmucosal neck (a) threaded body, (b) and active tip, (c) of mini-implants in the NeodentR (1), SINR (2), MorelliR (3), ConexaoR (4), and Foresta DentR (5) groups.

Despite the differences in the size of the orthodontic accessory on the screw head, they all had uniform structure and good surface finish. In Groups 1, 2, 3 and 4, the accessory was in the shape of an orthodontic button, whereas Group 5 mini-implants had bracket-shaped screw heads (Fig. 1). Mini-implants in Group 4 had a greater number of threads and flutes at the tip and a screw head diameter equal to that of the transmucosal neck. In Groups 1, 2, 3, and 5, the screw head and transmucosal neck had different diameters. Although all mini-implants were of the self-drilling variety, those in Groups 1, 3 and 5 had sharper tips (Fig. 1). Longitudinal sections were assessed to detect defects in the internal microstructure of each miniimplant, whereas cross-sections were compared against the ETTC-2 regarding the distribution of alpha and beta phases in the alloy. There were no visible imperfections in the inner structure of any mini-implants, and no internal defects were detected on the longitudinal sections. The mini-implants had a fine microstructure composed of an alpha matrix into which spheroidal beta-phase particles were dispersed. On cross sections, the internal microstructure of the alloys was consistent with ETTC-2 standard class A1. Alpha phase titanium appears light, whereas beta-phase granules appear darker. The small granule size of both phases and balanced alpha/beta ratio are indicative of high internal structure quality (Fig. 2).


Fig. 2: Photomicrograph showing the microstructure of the mini-implants (cross section)

DISCUSSION

We found significant differences in screw head, threaded body, and tip design between the five brands of mini-implants. Furthermore, SEM analysis showed surface irregularities and detritus, particularly at the implant tip. Metallographic analysis did not show any defects in the microstructure. All mini-implants tested met the European standard for titanium alloy production. Mini-implants are an effective and very well tolerated tool for skeletal anchorage, and have become the gold standard for orthodontic biomechanics in adults2. They are available in a variety of shapes, diameters, lengths, and titanium alloy compositions. However, it bears stressing that failure has been reported during mini-implant placement and removal. Mini-implant fractures are usually due to torsional strain caused by their small diameter7,15,21. Reicheneder et al.22 reported that different miniimplant systems showed comparable elementary composition. They stressed that differences in mechanical properties can be attributed to miniimplant design, and that implant morphology plays an essential role in ensuring primary stability.
In our study, the surface defects found in most samples, particularly at the active tip, may be caused by the machining process. These defects may be a starting point for electrochemical degradation processes that can alter the surface finish of the implant and its resistance and other material properties23. According to Sebbar et al.24 improvements in the surface treatment of mini-implants could improve their corrosion resistance. Mini-implants in Group 5 had fewer surface irregularities at the tip and better polish along the threaded body. The machining process determines the surface finish of the piece. Machining leads to a rough surface. Therefore, the biocompatibility of the surface texture has major influence on the type and progression of reactions in the tissues adjacent to the implant surface. Furthermore, changes in surface morphology that may occur during the sterilization process and mechanical damage that may be sustained during mini-implant placement and removal may induce changes in osteoblast growth and differentiation25,26. Studies have also analyzed changes in mini-implant design that might lead to improvements in the mechanical properties11. A greater number of threads and a finer pitch in the implant are associated with greater mechanical locking ability, enhanced resistance during mini-implant placement, improved resistance to displacement, and improved primary implant stability8. Mini-implants in Group 4 may improve the distribution of applied forces because they have a greater number of threads. Furthermore, the presence of flutes may be linked to greater fracture resistance, as it prevents concentration of excessive strain in the adjacent tissues8,26. Conversely, thread design may also interfere with the distribution of strain under load11. Hence, further studies are required to ascertain the influence of design on the mechanical properties of mini-implants.
Lee et al.27 found that many undesirable outcomes are attributable to the design of mini-implants. They claim that a coarsely finished or poorly designed mini-implant active tip may compromise final implant placement and primary stability. All miniimplants we tested were of the self-drilling variety, and those in Groups 1, 3 and 5 had the finest and sharpest tips, which suggests greater ease of placement without pilot hole drilling. The diameter of the mini-implant head is an important design factor. It should be wider than the transmucosal neck to prevent overgrowth of soft tissues. All implants had this design feature, except for those in Group 4. Casaglia et al.12 showed that small transmucosal neck diameter is a site of increased fragility. The authors detected microfissures and grooves on the surface and concluded that these irregularities may predispose to mini-implant fracture. Most mini-implants are made of Ti-6Al-4V (ASTM Grade 5 titanium alloy). This alloy has greater mechanical resistance than pure titanium and is more appropriate for small-diameter devices. Furthermore, its lower bioactivity facilitates implant removal because of less osseointegration18. Titanium alloys must be free of external irregularities and internal imperfections to avoid interference with fracture resistance, mechanical retention, displacement resistance, and primary stability16. Our metallographic analysis did not reveal any internal defects, corroborating the findings of Cotrim-Ferreira et al.15 and Eliades et al.16.
The alpha phase of titanium alloys is a soft alloy showing high resistance and tensile strength, but low ductility. Alpha-stabilizing elements increase the temperature range at which the alpha phase remains stable. The beta phase, in turn, has superior forming and fatigue resistance, but is highly vulnerable to atmospheric contamination. Beta stabilizers make the beta phase stable at low temperatures18. The matrix of all mini-implants assessed contained beta and alpha titanium, which indicates that the small amount of vanadium in the alloy was sufficient to retain significant amounts of beta-phase titanium, thus enhancing the properties of the alloy, as demonstrated by Iijima et al.14.
All mini-implants complied with the ETTC-2 standards, namely in class A1. Cotrim-Ferreira et al.15 found that the microstructure of ConexaoR brand implants was class A9, showing differences in hardness, resistance, and elastic modulus because of different alpha-beta phase ratio. However, this may not interfere with mechanical resistance, as alloys of any class from A1 through A10 can be used for mini-implant manufacturing according to the ETTC-2 standards. Despite the advantages of titanium alloy miniimplants, practitioners must be aware of the topographic and microstructural features of mini-implants, since they influence the effectiveness of orthodontic anchorage. Orthodontic treatment depends on reliable and effective anchorage. Primary stability, mechanical resistance, and clinical performance of miniimplants, in turn, depend on their topographical and microstructural characteristics.
The present study concluded that the orthodontic mini-implants assessed exhibited significant differences in the design of the screw head, transmucosal neck, threaded body, and active tip. Furthermore, surface irregularity and debris were found in all groups, particularly at the active tip. Conversely, no internal defects were detected, and all groups complied with the international standards for mini-implant manufacturing. Further studies of orthodontic mini-implants should prioritize topographic and microstructural analysis combined with mechanical testing.

REFERENCES

1. Labanauskaite B, Jankauskas G, Vasiliauskas A, Haffar N. Implants for orthodontic anchorage. Meta-analysis. Stomatologija 2005;7:128-132.         [ Links ]

2. Favero L, Brollo P, Bressan E. Orthodontic anchorage with specific fixtures: related study analysis. Am J Orthod Dentofacial Orthop 2002;122:84-94.         [ Links ]

3. Lim SA, Cha JY, Hwang CJ. Insertion torque of orthodontic miniscrews according to changes in shape, diameter and length. Angle Orthod 2008;78:234-240.         [ Links ]

4. Cheng SJ, Tseng IY, Lee JJ, Kok SH. A prospective study of the risk factors associated with failure of mini-implants used for orthodontic anchorage. Int J Oral Maxillofac Implants 2004;19:100-106.         [ Links ]

5. Florvaag B, Kneuertz P, Lazar F, Koebke J, Zoller JE, Braumann B, Mischkowski RA. Biomechanical properties of orthodontic miniscrews. An in-vitro study. J Orofac Orthop 2010;71:53-67.         [ Links ]

6. Mattos CT, Ruellas ACO, Elias CN. Is it possible to re-use mini-implants for orthodontic anchorage? Results of an in vitro study. Mater Res 2010;13:521-525. DOI: http://dx.doi.org/10.1590/S1516-14392010000400015        [ Links ]

7. Reynders R, Ronchi L, Bipat S. Mini-implants in orthodontics: a systematic review of the literature. Am J Orthod Dentofacial Orthop 2009;135:564.e1-19; discussion 564- 565.         [ Links ]

8. Brinley CL, Behrents R, Kim KB, Condoor S, Kyung HM, Buschang PH. Pitch and longitudinal fluting effects on the primary stability of miniscrew implants. Angle Orthod 2009;79:1156-1161.         [ Links ]

9. Park HS, Jeong SH, Kwon OW. Factors affecting the clinical success of screw implants used as orthodontic anchorage. Am J Orthod Dentofacial Orthop 2006;130:18-25.         [ Links ]

10. Kim TW, Baek SH, Kim JW, Chang YI. Effects of microgrooves on the success rate and soft tissue adaptation of orthodontic miniscrews. Angle Orthod 2008;78:1057- 1064.         [ Links ]

11. Chang JZ, Chen YJ, Tung YY, Chiang YY, Lai EH, Chen WP, Lin CP. Effects of thread depth, taper shape, and taper length on the mechanical properties of mini-implants. Am J Orthod Dentofacial Orthop 2012;141:279-288.         [ Links ]

12. Casaglia A, Dominici F, Pachi F, Turla R, Cerroni L. Morphological observations and fractological considerations on orthodontics miniscrews. Minerva Stomatol 2010;59: 465-476.         [ Links ]

13. Song YY, Cha JY, Hwang CJ. Mechanical characteristics of various orthodontic mini-screws in relation to artificial cortical bone thickness. Angle Orthod 2007;77: 979-985.         [ Links ]

14. Iijima M, Muguruma T, Brantley WA, Okayama M, Yuasa T, Mizoguchi I. Torsional properties and microstructures of miniscrew implants. Am J Orthod Dentofacial Orthop 2008;134: 333.e1-6; discussion 333-4.         [ Links ]

15. Cotrim-Ferreira FA, Quaglio CL, Peralta RP, Carvalho PE, Siqueira DF. Metallographic analysis of the internal microstructure of orthodontic mini-implants. Braz Oral Res 2010;24:438-442.         [ Links ]

16. Eliades T, Zinelis S, Papadopoulos MA, Eliades G. Characterization of retrieved orthodontic miniscrew implants. Am J Orthod Dentofacial Orthop 2009;135:10.e1-7; discussion 10-11.         [ Links ]

17. Donachie M. Titanium: a technical guide. 2nd ed. Ohio: ASM International, 2000: 381 p.         [ Links ]

18. American Society for Testing and Materials. ASTM E3-01 Standard guide for preparation of metallographic specimens. West Conshohocken: American Society for Testing and Materials 2012: 13 p.         [ Links ]

19. American Society for Testing and Materials. ASTM E407- 99 Standard practice for nicroetching netals and alloys. West Conshohocken: American Society for Testing and Materials, 1999. 21 p.         [ Links ]

20. American Society for Testing and Materials. ASTM E7-03 Standard terminology relating to metallography. West Conshohocken: American Society for Testing and Materials, 2009. 30 p.         [ Links ]

21. Papadopoulos MA, Tarawneh F. The use of miniscrew implants for temporary skeletal anchorage in orthodontics: a comprehensive review. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 2007; 103: e6-15.         [ Links ]

22. Reicheneder C, Rottner K, Bokan I, Mai R, Lauer G, Richter G, Gedrange T, Proff P. Mechanical loading of orthodontic miniscrews - significance and problems: an experimental study. Biomed Tech 2008; 53:242-245.         [ Links ]

23. Morais LS, Serra GG, Muller CA, Andrade LR, Palermo EF, Elias CN, Meyers M. Titanium alloy mini-implants for orthodontic anchorage: immediate loading and metal ion release. Acta Biomater 2007;3:331-339.         [ Links ]

24. Sebbar M, Bourzgui F, Aazzab B, Elquars F. Anchorage miniscrews: a surface characterization study using optical microscopy. Int Orthod 2011; 9:325-338.         [ Links ]

25. Cooper LF. Biologic determinants of bone formation for osseointegration: clues for future clinical improvements. J Prosthet Dent 1998; 80:439-449.         [ Links ]

26. Schwartz Z, Lohmann CH, Blau G, Blanchard CR, Soskolne AW, Liu Y, Cochran DL, Dean DD, Boyan BD. Re-use of implant coverscrews changes their surface properties but not clinical outcome. Clin Oral Implants Res 2000; 11:183-194.         [ Links ]

27. Lee JS, Kim JK, Park Y, Vanarsdall RL. Application of orthodontic mini-implants. Chicago: Quintessence, 2007. 286 p.         [ Links ]

Creative Commons License All the contents of this journal, except where otherwise noted, is licensed under a Creative Commons Attribution License